This report describes research efforts to provide practitioners with 1) well-defined performance specifications to fairly evaluate and compare instruments; 2) procedures on how to validate that an instrument meets specifications upon installation; 3) quality assurance procedures to verify ongoing instrument performance; and 4) a custom set of physical reference standards designed specifically for calibration and quality control of surface topography instruments being used for firearm and toolmark identification.
The emerging capability for measuring and analyzing the three-dimensional (3D) surface topography of forensic samples is expected to have a profound impact on firearm and toolmark identification. As forensic laboratories are working to integrate 3D topography metrology into actual casework, it is critical that quality assurance guidance is in place on how to select a suitable measurement instrument and how to ensure that the instrument performs to specification. Although progress towards these goals is being made in several standards organizations for generic applications, there remains a need for additional detailed documentation that applies to firearm and toolmark identification. (publisher abstract modified)
Downloads
Similar Publications
- A Case Study Evaluating the Efficacy of an Ad Hoc Hospital Collection Device for Fentanyl in Infant Oral Fluid
- FRVT 2006 and ICE 2006 Large-Scale Experimental Results
- Forced Condensation of Cyanoacrylate with Temperature Control of the Evidence Surface to Modify Polymer Formation and Improve Fingerprint Visualization