This project compared single quadruple mass spectrometry and diode array-ultraviolet (PDA-UV) detection interfaced to ultra-high performance supercritical fluid chromatography for the quantitative analysis of synthetic cathinones.
Synthetic cathinones, also known as bath salts, are derived from cathinone, a component of the khat plant. For 15 controlled solutes linearity, repeatability, and limits of detection were determined using both UV and MS detection. Quantitation studies were performed, using the above detectors for 19 mixtures of up to 4 of the above compounds with an adulterant to simulate seized samples. MS detection provided approximately an order of magnitude greater linearity range and allowed for two to three orders of magnitude lower limits of detection than UV detection. Both detection techniques exhibited similar results of analysis and comparable repeatability. The latter detection mode which provided significantly high linearity correlation coefficients would be preferred for quantitative analysis. (publisher abstract modified)
Downloads
Similar Publications
- Comparative Evaluation of Genotyping Technologies for Investigative Genetic Genealogy in Sexual Assault Casework
- Elucidation of the Effect of Solar Light on the Near-Infrared Excitation Raman Spectroscopy-Based Analysis of Fabric Dyes
- Dyed Hair and Swimming Pools: The Influence of Chlorinated and Nonchlorinated Agitated Water on Surface-Enhanced Raman Spectroscopic Analysis of Artificial Dyes on Hair