The quantification of elemental impurities in aluminum oxide powder is important due to its large demand for various industrial applications. One of the techniques to perform such analysis is Laser-Induced Breakdown Spectroscopy (LIBS) and to that end, powders are usually pressed into pellets; however, the density of the powder-compact is often not reported and its effect on the analytical figure-of-merit is unknown. The results of the current study indicate that the sensitivity increases when the grain size increases for a given density and that its relative standard deviation decreases with the density of the powder compacts without any grain size effect. These results imply that both the powder-compact density and the ultimate particle size of the powders must be specified in the protocol for building the calibration curve. (publisher abstract modified)
Downloads
Similar Publications
- Rapid DNA Crime Scene Sample Multi-Laboratory Evaluation Project
- Physical and Biochemical Factors Affecting the Recovery and Analysis of DNA from Human Skeletal Remains
- Audit of the Office of Justice Programs Victim Compensation Grants Awarded to the Florida Department of Legal Affairs, Tallahassee, Florida