U.S. flag

An official website of the United States government, Department of Justice.

NCJRS Virtual Library

The Virtual Library houses over 235,000 criminal justice resources, including all known OJP works.
Click here to search the NCJRS Virtual Library

Attenuated Total Reflection (ATR) Sampling in Infrared Spectroscopy of Heterogeneous Materials Requires Reproducible Pressure Control

NCJ Number
252315
Journal
Applied Spectroscopy Volume: 71 Issue: 1 Dated: 2017 Pages: 97-104
Author(s)
Zhenyu Lu; Brianna M. Cassidy; Stephanie A. Dejong; Raymond G. Belliveau; Michael L. Myrick; Stephen L. Morgan
Date Published
2017
Length
8 pages
Annotation
The authors of this article report that while collecting data from fabrics with heterogeneous coatings using infrared spectroscopy, they observed systematic pressure-dependent changes in spectra that can be eliminated by more reproducible pressure control; and they also describe a pressure sensor adapted to work with an attenuated total reflection (ATR) tower to enable more precise control of pressure during ATR sampling.
Abstract
Attenuated total reflection Fourier transform infrared (ATR FT-IR) spectroscopy, in which the sample is pressed against an internal reflection element, is a popular technique for rapid IR spectral collection; however, depending on the accessory design, the pressure applied to the sample is not always well controlled. (Publisher abstract modified)