U.S. flag

An official website of the United States government, Department of Justice.

NCJRS Virtual Library

The Virtual Library houses over 235,000 criminal justice resources, including all known OJP works.
Click here to search the NCJRS Virtual Library

Simultaneous colorimetric detection of metallic salts contained in low explosives residue using a microfluidic paper-based analytical device (µPAD)

NCJ Number
304487
Journal
Forensic Chemistry Volume: 9 Dated: June 2018 Pages: 35-41
Author(s)
Kathryn R. Chabaud ; Jennifer L. Thomas; Michelle N. Torres; Sheila Oliveira ; Bruce R .McCord
Date Published
June 2018
Length
7 pages
Annotation

This article reports on the development of a microfluidic paper-based analytical device (µPAD) that has been developed for the onsite determination of low explosives residues.

Abstract

The device is capable of simultaneous detection of a variety of metallic salts using colorimetric detection. The µPAD was printed on chromatography paper with wax ink, allowing for the creation of a set of hydrophobic channels. Each channel contains a specific set of reagents that yields a color change upon interaction with a specific metal salt. The device is capable of performing six simultaneous tests, including the detection of metallic salts present in primer residues and pyrotechnic low explosive devices. Metals detected include lead, barium, antimony, iron, aluminum, zinc, and magnesium. Detection times were found to be less than 10 min and visual limits of detection ranged from 0.025 to 0.4 µg of these metallic compounds. The resultant paper chip was then tested against various interferences and used to examine a number of different pyrotechnic compositions. This new device should prove useful in onsite detection of post-blast residues of pyrotechnics in the field due its portability and ease of use. (Publisher Abstract)